Your nose cannot tell you a manhole is deadly. Oxygen-deficient air has no smell, hydrogen sulfide paralyzes your sense of smell at the very concentrations that kill, and flammable vapors can sit just below the explosive range with no warning. The only thing between you and the atmosphere is a gas monitor, and a monitor is only trustworthy if it is bump tested, calibrated, set to the right alarms and used correctly. This guide teaches the instruments, the testing order OSHA requires, bump test versus calibration, documentation, and how the same calibration discipline applies to flow and blower meters in collection work.
1. What the regulations require of the instrument
OSHA's permit-required confined space standard, 29 CFR 1910.146, sets the basics:
- Test before entry to determine whether acceptable entry conditions exist, and test or monitor as necessary while entry continues